Safety and Tolerability as Measured by Adverse Events, Local Tolerability of Infusion Site Reactions, and Clinical Laboratory Parameters
| Group | Value (Participants) |
|---|---|
| Part 1, ZP4207 | 8 |
| Part 1, Lilly Glucagon | 12 |
Trial · Interventional
CompletedPhase 2Results postedRegistry dataShort name dasiglucagon
United States
1 trial · 1 completed · 1 with posted results
Text and fields below are the sponsor's entries at ClinicalTrials.gov, shown as posted.
The purpose of this study was to determine whether the Bionic Pancreas with ZP4207 (dasiglucagon\*) was feasible to improve glycemic control in adults with type 1 diabetes mellitus. \*dasiglucagon is the proposed International Nonproprietary Name (pINN) for ZP4207
Experimental
Part 1, Lilly glucagon then ZP4207
In Part 1, 12 patients participated in 1-day treatment arms in random order (iPhone-based Bionic Pancreas using Lilly glucagon and iPhone-based Bionic Pancreas using ZP4207 (dasiglucagon) {experimental drug} with insulin lispro) according to pre-generated randomization scheme.
Experimental
Part 1, ZP4207 then Lilly Glucagon
In Part 1, 12 patients participated in 1-day treatment arms in random order (iPhone-based Bionic Pancreas using Lilly glucagon and iPhone-based Bionic Pancreas using ZP4207 (dasiglucagon) {experimental drug} with insulin lispro) according to pre-generated randomization scheme.
Experimental
Part 2, Lilly glucagon then ZP4207
In Part 2, it was planned to enrol up to 10 new patients to participate in 1-day treatment arms in random order (iLet-based Bionic Pancreas using Lilly glucagon and iLet-based Bionic Pancreas using ZP4207 (dasiglucagon) {experimental drug} with insulin lispro) according to pre-generated randomization scheme. However, due to unavailability of the iLet, the sponsor decided to stop the trial upon completion of Part 1. Part 2 of the trial using the iLet was consequently not conducted.
Experimental
Part 2, ZP4207 then Lilly Glucagon
In Part 2, it was planned to enrol up to 10 new patients to participate in 1-day treatment arms in random order (iLet-based Bionic Pancreas using Lilly glucagon and iLet-based Bionic Pancreas using ZP4207 (dasiglucagon) {experimental drug} with insulin lispro) according to pre-generated randomization scheme. However, due to unavailability of the iLet, the sponsor decided to stop the trial upon completion of Part 1. Part 2 of the trial using the iLet was consequently not conducted.
Outcomes the sponsor listed as primary.
Safety and Tolerability as Measured by Adverse Events, Local Tolerability of Infusion Site Reactions, and Clinical Laboratory Parameters
Time frame · Up to 50 days
Safety and tolerability of ZP4207 in the BP using either the iPhone or the iLet platform, as measured by adverse events (AEs), local tolerability of infusion site reactions, and clinical laboratory parameters. See adverse events section for results on AEs by system organ class and preferred term. Clinical laboratory parameters in terms of overall 'investigations' AEs and abnormal hematology parameters that did not resolve by the follow-up visit are presented below. LLN = lower limit of the normal range. Investigations and vital signs AEs by preferred term are presented in the AE section. Participants with infusion site pain and nausea measured by visual analog scales (VAS) are presented below; mean values are presented under secondary outcomes. For the VAS, individuals marked on a 10-cm line corresponding to the amount of pain or nausea being experienced, with low scores (cm) indicating no feelings of pain or nausea and high scores (cm) indicating high feelings of pain or nausea.
Pain Measured on a Visual Analog Scale (VAS)
Time frame · 16 hours
The VAS scale was used to measure pain at the end of the visit (16 hours) for patients in both treatment groups. The VAS was a psychometric response scale used to measure subjective characteristics of pain. Patients marked a location on a 0 to 10-cm line that corresponded to the amount of pain being experienced, with low scores (cm) indicating no feelings of pain and high scores (cm) indicating high feelings of pain. Actual values are shown. The maximum value in the Lilly glucagon group was recorded at hour 3.
Nausea Measured on a Visual Analog Scale (VAS)
Time frame · 16 hours
The VAS scale was used to measure nausea at the end of the visit (16 hours) for patients in both treatment groups. The VAS was a psychometric response scale used to measure subjective characteristics of nausea. Patients marked a location on a 0 to 10-cm line that corresponded to the amount of nausea being experienced, with low scores (cm) indicating no feelings of nausea and high scores (cm) indicating high feelings of nausea. Actual values are shown. The maximum values in both groups were recorded at hour 6, the start of the exercise period.
Glycemic Regulation
Time frame · 16 hours
Measure glycemic regulation, including hypoglycemia exposure (percent of time spent with continuous glucose monitor \[CGM\] glucose\<60mg/dL)
Average Percent Glucagon Dose Amounts Calculated by the Bionic Pancreas Control Algorithm That Are Successfully Delivered by the Pump.
Time frame · 16 hours
Secondary endpoint of bionic pancreas function, presented by treatment group. The analysis of bionic pancreas function endpoints was on an intention-to-treat basis.
Average Percent Insulin Dose Amounts Calculated by the Bionic Pancreas Control Algorithm That Are Successfully Delivered by the Pump.
Time frame · 16 hours
Secondary endpoint of bionic pancreas function, presented by treatment group. The analysis of bionic pancreas function endpoints was on an intention-to-treat basis.
Average Percentage of Time During Which the Bionic Pancreas is Functioning Nominally in All Respects Based on Real-time Continuous Glucose Monitoring (CGM) Data
Time frame · 16 hours
Secondary endpoint of bionic pancreas function, presented by treatment group. The analysis of bionic pancreas function endpoints was on an intention-to-treat basis.
Average Percentage of Time During Which the Bionic Pancreas is Functioning Nominally With or Without a New CGM Glucose Reading Captured
Time frame · 16 hours
Secondary endpoint of bionic pancreas function, presented by treatment group. The analysis of bionic pancreas function endpoints was on an intention-to-treat basis.
CGM Reliability Index, Calculated as Percentage of Possible Values Actually Recorded by CGM
Time frame · 16 hours
Secondary endpoint of bionic pancreas function, presented by treatment group. The analysis of bionic pancreas function endpoints was on an intention-to-treat basis.
CGM Mean Absolute Relative Difference Versus Time-stamped Blood Glucose (BG) Values From Meter Download
Time frame · 16 hours
Secondary endpoint of bionic pancreas function, presented by treatment group. The analysis of bionic pancreas function endpoints was on an intention-to-treat basis.
Number of Patients With Technical Faults Associated With the BP Including Cause and Resolution: Calibration Issues
Time frame · 16 hours
Technical faults in terms of calibration issues were listed by patient.
Number of Patients With Technical Faults Associated With the BP Including Cause and Resolution: Connectivity Issues
Time frame · 16 hours
Technical faults related to connectivity issues were listed
Diabetes Treatment Satisfaction Questionnaire - Status
Time frame · Up to 3 months
This questionnaire was not assessed as per protocol amendment 7.
Diabetes Treatment Satisfaction Questionnaire - Change
Time frame · Up to 3 months
This questionnaire was not assessed as per protocol amendment 7.
T1-Diabetes Distress Scale
Time frame · Up to 3 months
This questionnaire was not assessed as per protocol amendment 7.
Problem Areas in Diabetes Survey
Time frame · Up to 3 months
This questionnaire was not assessed as per protocol amendment 7.
Hypoglycemia Fear Survey
Time frame · Up to 3 months
This questionnaire was not assessed as per protocol amendment 7.
Impact of Daily Diabetes Demands
Time frame · Up to 3 months
This questionnaire was not assessed as per protocol amendment 7.
Bionic Pancreas User Opinion Survey
Time frame · Up to 3 months
This questionnaire was not assessed as per protocol amendment 7.
Values are shown as posted at ClinicalTrials.gov, with their units. PepBank has not converted, pooled or compared them.
| Group | Value (Participants) |
|---|---|
| Part 1, ZP4207 | 8 |
| Part 1, Lilly Glucagon | 12 |
| Group | Value (cm) | Spread |
|---|---|---|
| Part 1, ZP4207 | 0.00 | 0.00 |
| Part 1, Lilly Glucagon | 0.01 | 0.0029 |
| Group | Value (cm) | Spread |
|---|---|---|
| Part 1, ZP4207 | 0.00 | 0.00 |
| Part 1, Lilly Glucagon | 0.00 | 0.00 |
| Group | Value (percentage of time points) | Spread |
|---|---|---|
| Part 1, ZP4207 | 12.78 | 17.262 |
| Part 1, Lilly Glucagon | 17.60 | 14.080 |
| Group | Value (Percentage of treatment delivered) | Spread |
|---|---|---|
| Part 1, ZP4207 | 96.73 | 4.021 |
| Part 1, Lilly Glucagon | 97.01 | 2.851 |
| Group | Value (Percentage of treatment delivered) | Spread |
|---|---|---|
| Part 1, ZP4207 | 98.36 | 1.114 |
| Part 1, Lilly Glucagon | 97.73 | 2.051 |
| Group | Value (percentage of time functioning nominally) | Spread |
|---|---|---|
| Part 1, ZP4207 | 95.41 | 1.846 |
| Part 1, Lilly Glucagon | 94.26 | 3.966 |
| Group | Value (percentage of time functioning nominally) | Spread |
|---|---|---|
| Part 1, ZP4207 | 97.86 | 1.125 |
| Part 1, Lilly Glucagon | 96.31 | 3.891 |
| Group | Deaths | Serious events | Other events |
|---|---|---|---|
| Part 1, ZP4207In Part 1, 10 patients received treatment with ZP4207 (dasiglucagon) {experimental drug} with insulin lispro in the iPhone-based Bionic Pancreas according to a pre-generated randomization scheme. Insulin Lispro: Used to lower blood glucose. Commercially available by prescription and is indicated for patients with T1DM, but not for use in a bionic pancreas. Individualized dose based on metabolic needs and frequent monitoring of blood glucose. ZP4207 (dasiglucagon): A glucagon analog not yet approved by the FDA. Subcutaneous administration in one BP arm. iPhone-based bionic pancreas: An experimental device. | 0 / 10 | 0 / 10 | 8 / 10 |
| Part 1, Lilly GlucagonIn Part 1, 12 patients received treatment with Lilly glucagon and insulin lispro in the iPhone-based Bionic Pancreas according to a pre-generated randomization scheme. Insulin Lispro: Used to lower blood glucose. Commercially available by prescription and is indicated for patients with T1DM, but not for use in a bionic pancreas. Individualized dose based on metabolic needs and frequent monitoring of blood glucose. Glucagon: A hormone normally made by the pancreas to raise blood glucose. Used to treat low blood sugar. Commercially available by prescription and is indicated for patients with T1DM in severe hypoglycemia, but not for use in a BP. Subcutaneous administration in one BP arm. iPhone-based bionic pancreas: An experimental device. | 0 / 12 | 0 / 12 | 12 / 12 |
Each cell reads participants affected / participants at risk, as the registry posted them.
Registry data comes from ClinicalTrials.gov. Study results come from studies, not from the registry. How PepBank works