Percentage of Time in Target Range Over 24 Hour
| Group | Value (percentage of time) |
|---|---|
| Bihormonal Closed Loop | 2 |
| Standard Opened Loop Pump | 4.7 |
| Insulin Only | 1.3 |
Trial · Interventional
CompletedNot applicableResults postedRegistry dataThis trial does not test a peptide that PepBank tracks.
United Kingdom
1 trial · 1 completed · 1 with posted results
Clinical Assessment of a Closed-loop System With Glucagon, Exercise and Mixed Meals
NCT023972653 Dec 2014 to 22 Jul 2015
Text and fields below are the sponsor's entries at ClinicalTrials.gov, shown as posted.
The diabetes technology group at Imperial College have developed a bio-inspired artificial pancreas (BiAP) system which uses a control algorithm based on a mathematical model of beta-cell physiology. The algorithm is implemented on a miniature silicon microchip within a portable handheld device, which interfaces the components of the artificial pancreas. Development of closed-loop insulin delivery devices to intensify control without hypoglycaemia has been extensively reviewed and have shown encouraging results . However, they have not yet proven to be robust when challenged with uncertainty and the external challenges (such as mixed meal contents, physical exercise, physiological stress and intercurrent illness) that people with Type 1 Diabetes Mellitus (T1DM) may be exposed to outside the clinical environment. The principal research objective is to assess the safety and efficacy of a closed-loop system for T1DM compared to standard insulin pump therapy (open-loop). The primary outcome from the studies will be % time spent with a glucose concentration in the target range (3.9-10.0mmol/l). This outcome incorporates safety as it ensures subjects do not have low or high glucose excursions and is the principal measure of efficacy for closed-loop insulin delivery systems in the scientific literature. Other measured outcomes will be % time spent in euglycaemia (3.9-7.8mmol/l), % time spent in hypoglycaemia (\<3.9mmol/l), % time spent in hyperglycaemia (\>10mmol/l), mean venous blood and sensor glucose, glycaemic variability as measured by standard metrics (Standard Deviation, Continuous Overlapping Net Glycaemic Action, Lability Index, J-Index, Glycaemic Risk Assessment Diabetes Equation, Mean Of Daily Differences, Mean Amplitude of Glucose Excursion, Average Daily Risk Range, M-VALUE, Mean Average Glucose), glycaemic risk as measured by Low Blood Glucose Index (LBGI) and High Blood Glucose Index (HBGI), closed-loop error grid analysis, glucose area under the curve. All measures have been previously published and validated. This clinical trial protocol assesses the artificial pancreas system in three separate sub-studies: 1. In a bi-hormonal (insulin and glucagon) configuration 2. During and after exercise with bi-hormonal closed loop, and standard insulin opened loop 3. During and after meals of mixed composition with bi-hormonal closed loop, and standard insulin opened loop
Experimental
Bihormonal closed loop
Bi-hormonal closed loop pump will be used in the exercise and mixed meal substudies
Active comparator
Standard opened loop pump
Standard opened loop pump will be used in the exercise and mixed meal substudies
Placebo comparator
Insulin only
Insulin only closed loop
Outcomes the sponsor listed as primary.
Percentage of Time in Target Range Over 24 Hour
Time frame · 24 hours
The primary outcome from the studies will be time spent with a glucose concentration in the target range (3.9-10.0mmol/l).
Values are shown as posted at ClinicalTrials.gov, with their units. PepBank has not converted, pooled or compared them.
| Group | Value (percentage of time) |
|---|---|
| Bihormonal Closed Loop | 2 |
| Standard Opened Loop Pump | 4.7 |
| Insulin Only | 1.3 |
| Group | Deaths | Serious events | Other events |
|---|---|---|---|
| Bihormonal Closed LoopBi-hormonal closed loop pump will be used in the exercise and mixed meal substudies Bi-hormonal closed loop pump: Using a bio-inspired artificial pancreas consisting of a bi-hormonal closed loop pump | 0 / 6 | 0 / 6 | 0 / 6 |
| Standard Opened Loop PumpStandard opened loop pump will be used in the exercise and mixed meal substudies Standard opened loop pump: Using a standard opened loop pump | 0 / 6 | 0 / 6 | 0 / 6 |
| Insulin OnlyInsulin only closed loop | 0 / 6 | 0 / 6 | 0 / 6 |
Each cell reads participants affected / participants at risk, as the registry posted them.
Registry data comes from ClinicalTrials.gov. Study results come from studies, not from the registry. How PepBank works
Studies that cite this trial.
1 study on this page
PubMed tag: none · 2013
Herrero et al. 2013Journal of diabetes science and technologycited by 16
PMID 23911175PMC3879758DOI 10.1177/193229681300700416PUB-23911175
Who could join, as posted.
PepBank is a research reference. Talk to a doctor before you use any peptide.
In PepBank now:147 peptides, 6 blends and 13 other compounds97,577 study results from 17,383 studies8,473 clinical trialsUpdated