In HPLC-RIA of DSIP, carry-over (shadowing) from the previous standard was as high as 10%, which can cause false positive results.
In plain wordsIn lab tests, up to 10% of DSIP carried over between HPLC runs, which can give false positive results.
- Outcome
- Not applicableHPLC shadowing of DSIP
- Study
n: not reported in source
HPLC with radioimmunoassay- Outcome group
- Analytical methods and detection · Pharmacology
Measurements
- HPLC shadowing
- ≤ 10%
- DSIP
Source sentences
With delta sleep-inducing peptide (DSIP), we found the shadowing to be as high as 10%, although it was only 1% with 125I-Tyr-DSIP.
1 context sentence(design, population, supporting numbers)
The shadow phenomenon, representing carry-over of the peptide from previous application of the standard, can appear to result in the presence of endogenous peptide in the test sample when none is actually there.
In this study, Delta sleep-inducing peptide (DSIP) was a lab tool or a carrier, part of a mix, or the subject of a lab method. This result is not evidence about taking Delta sleep-inducing peptide (DSIP).